Masthead
 

Answer to Question #6789 Submitted to "Ask the Experts"

Category: Instrumentation and Measurements — Instrument Calibration (IC)

The following question was answered by an expert in the appropriate field:

Q

How can I calibrate an RIA (radioimmunoassay) gamma counter for measuring 125I? Which protocol shall I use? What are the standard-reference sources that I have to use? How often do I use them? Are there any documents that I can read in order to start such a procedure?


A

I am not familiar with any specific protocols recommended by government or other standards-setting or recognized recommending groups for calibrating gamma counters. There are other protocols and reports that relate to calibration of systems, such as scintillation detectors with analyzers used for gamma energy analysis, such as the 2007 Standard IEC 61453 of the International Electrotechnical Commission that might be of interest to you. I will attempt to address some of the considerations that do apply to calibration of the gamma counter.

You can calibrate the gamma counter for counting 125I using an appropriate standard of 125I. The standard should be prepared in a geometry and composition as similar as possible to the samples that you will be counting. The iodine emits rather low-energy photons in the interval from about 27 keV to 35 keV, and the tube or container used to hold the standard should also be the same as that used in counting so as not to produce wall attenuation effects that are different from that of the sample. You may be able to get the supplier to make up the standard according to your needs. If not, you may have to purchase the standard as a calibrated solution and use a portion of it to make up the standard in the necessary form. The activity of the standard should be traceable to an accepted standards laboratory, such as the National Institute of Standards and Technology or a secondary laboratory.

The instrument must be set up so that the major 125I counts will fall in the counting window specified for 125I. Once this has been done, count the standard long enough to obtain enough counts to obtain an acceptable error—e.g., neglecting background, 100,000 counts yields a relative standard deviation of 0.3 percent. The amount of activity commonly used for calibration in this type of system is usually between 0.05 microcuries and 0.1 microcuries. This is usually sufficient to produce a statistically valid number of counts in a one-minute counting interval without producing high-enough count rates to saturate the system. It is generally desirable to take a series of consecutive counts that can be used to generate the mean count and, perhaps, to perform one or more appropriate performance tests (see below).

At the time that you do the initial calibration, it is also a good idea to perform a standard statistical test to validate the operation of the counting system. You can do this using the standard or an alternative source if you prefer. (Using the standard has the advantage that you will be able to use the data to generate a good value for the mean count that can be used in your efficiency determination). A Chi-square (χ2) test is one of the most common and useful tests in this regard. I would suggest taking 20 consecutive one-minute counts to perform this test. The value of χ2 is given by

n
          χ2 = Σ(xi - xm)2/xm,
i=1

where xi is the ith count, xm is the mean of all the n counts, and the summation of the squared differences from the mean is for all of the counts. The ideal value of χ2 is equal to the number of degrees of freedom for the data set (for this case the number of degrees of freedom is n-1 or 19 for a set of 20 counts). You can review a table of the values of χ2 in order to specify a range of acceptable values consistent with your level of acceptance (for more discussion see Question 5982 and its answer on the Health Physics Society Ask the Experts Web site). For example, for a rejection level of 5 percent (i.e., a willingness to conclude 5 percent of the time that good data are not acceptable), the range of acceptable χ2 values is from 8.91 to 32.7, and any value between these two would indicate acceptable performance of the counter.

When you do the initial calibration you might also want to count a source of 129I, a radionuclide often used as a stand-in for 125I for routine operational checks and secondary calibrations. The 129I has a very long half-life, and using it for future instrument performance checks and calibrations is often desirable since the 125I standard will decrease in activity appreciably with time and soon require replacement. The 129I is often referred to as mock 125I because it emits photons that are quite similar in energies to those from 125I and provides a reasonable simulation of the 125I.

The 129I source should be used to test the operation and consistency of performance of the gamma counter every day that the instrument is used. This is part of the routine quality-assurance program that should be in operation. If time permits, I would recommend performing a series of 10 or 20 counts each day and performing a Chi-square test of the counter. The frequency of calibration varies with the kind of use that the instrument gets. If it is being used heavily by multiple users, then calibrations should be performed more frequently than for cases involving one or two users and/or light use. Calibration frequencies may vary from about weekly to annually. Here is a link to a service manual for the Isocomp I gamma counter from MGM Instruments that describes some statistical recommendations and shows typical data that might be obtained in the calibration process, which is recommended on a weekly basis.

Since the calibration process is quite straightforward and requires rather little time to accomplish, once the initial calibration with the 125I has been done, I would feel comfortable using the 129I source on at least a monthly basis to perform a secondary calibration. In any case, be sure that your daily checks with the 129I source are conducted to confirm consistent performance. Naturally, if other radionuclides are also being assessed with the gamma counter, you would have to perform calibrations for them as well. I hope this is of some help to you.

George Chabot, PhD, CHP

Answer posted on 19 September 2007. The information and material posted on this Web site is intended as general reference information only. Specific facts and circumstances may alter the concepts and applications of materials and information described herein. The information provided is not a substitute for professional advice and should not be relied upon in the absence of such professional advice specific to whatever facts and circumstances are presented in any given situation. Answers are correct at the time they are posted on the Web site. Be advised that over time, some requirements could change, new data could be made available, or Internet links could change. For answers that have been posted for several months or longer, please check the current status of the posted information prior to using the responses for specific applications.